• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-360; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problem Suction Failure (4039)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Event Description
A customer reported error message "2016 bf probe suction failure" on the aia-360 analyzer.The customer primed the wash but error persisted.The customer states fluid was seen on top of the carousel near the wash probe well and several beads was seen in the well.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for follicle stimulating hormone (fsh), cardiac troponin i (ctnl2), luteinizing hormone (lh ii), prolactin (prl), and progesterone (prog iii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs and reproduced error by performing daily checks and primes.Fse resolved complaint by replacing the waste pump.Fse cleaned and lubricated the guide rails, validated analyzer by performing background checks and quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 operator's manual under chapter 7, section 7.1: list of error messages states the following: [2016] bf probe suction failure is generated when the suction by the bf probe is abnormal.The operator is instructed to contact the service department.The most probable cause of the reported event is due to failure of the waste pump.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key16075882
MDR Text Key308489513
Report Number3004529019-2022-00219
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/29/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-