A customer reported error message "2016 bf probe suction failure" on the aia-360 analyzer.The customer primed the wash but error persisted.The customer states fluid was seen on top of the carousel near the wash probe well and several beads was seen in the well.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for follicle stimulating hormone (fsh), cardiac troponin i (ctnl2), luteinizing hormone (lh ii), prolactin (prl), and progesterone (prog iii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
A field service engineering (fse) was at the customer's site to address reported event.Fse confirmed the complaint by reviewing the error logs and reproduced error by performing daily checks and primes.Fse resolved complaint by replacing the waste pump.Fse cleaned and lubricated the guide rails, validated analyzer by performing background checks and quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The aia-360 operator's manual under chapter 7, section 7.1: list of error messages states the following: [2016] bf probe suction failure is generated when the suction by the bf probe is abnormal.The operator is instructed to contact the service department.The most probable cause of the reported event is due to failure of the waste pump.
|