• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNKNOWN ROTAREX; THROMBECTOMY & ATHERECTOMY Back to Search Results
Catalog Number 80202
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/02/2022
Event Type  malfunction  
Event Description
It was reported that during a recanalization procedure in the distal lesion via middle segment of the superficial femoral artery approach, the coil spring in the catheter was allegedly ruptured.It was further reported that the balloon was used for dilatation.The procedure was completed placing another stent.There was no reported patient injury.
 
Manufacturer Narrative
The medical device manufacturer and manufacturing location for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.The catalog number identified has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in common device name and pma/510k.H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, photos were provided for review.The investigation of the reported event is currently underway.(expiration date: 01/2025).
 
Event Description
It was reported that during a recanalization procedure in the distal lesion via middle segment of the superficial femoral artery approach, the coil spring in the catheter was allegedly ruptured.It was further reported that the balloon was used for dilatation.The procedure was completed placing another stent.There was no reported patient injury.
 
Manufacturer Narrative
H10: the medical device manufacturer (d3) and manufacturing location (g1) for the straub product was selected as unknown due to system limitations.The correct medical device manufacturer and manufacturing location are straub medical us.H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the rotarex products that are cleared in the us.The pro code and 510 k number for the rotarex products are identified in d2 and g4.H10: manufacturing review: a manufacturing review was conducted and there was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was returned for evaluation and a physical investigation was performed for the catheter.The helix was broken at 20 cm distance from the tip of the catheter.Therefore, the investigation is confirmed for the reported break issue.The definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Uacy.H10: d4 (expiration date: 01/2025), g3 h11: h6 (method, result, conclusion) h11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAREX
Type of Device
THROMBECTOMY & ATHERECTOMY
Manufacturer (Section D)
UNKNOWN
BR 
Manufacturer (Section G)
UNKNOWN
BR  
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16076178
MDR Text Key308368292
Report Number3008439199-2022-00197
Device Sequence Number1
Product Code MCW
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K211738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number80202
Device Lot Number220314
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/02/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-