Model Number D134805
Device Problem Patient Device Interaction Problem (4001)
Patient Problem Cardiac Tamponade (2226)
Event Date 12/02/2022
Event Type  Injury  
Manufacturer Narrative
The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturing record evaluation (mre) cannot be conducted because no lot number was provided by the customer.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an idvt ¿ left ablation procedure with a thermocool® smart touch® sf bi-directional navigation catheter.The patient suffered cardiac tamponade requiring a pericardiocentesis.The patient suffered a pericardial effusion.Towards the end of the case, the physician was using intracardiac echocardiography (ice) to check his catheter position and noticed the effusion.A transthoracic echo was performed to confirm the pericardial effusion.A pericardiocentesis was performed and about 15 cc of fluid was removed from the pericardial space.Patient is stable currently and was stable throughout the entire procedure.No ablation was performed as there was no area to ablate.The physician thinks that the effusion happened from the transseptal using the agilis sheath.They are unsure if the catheter has been preserved for return and noted that the packaging has been discarded so the lot is unknown.Additional information was received.The adverse event was discovered during use of biosense webster products.Physician¿s opinion on the cause of this adverse event was transseptal procedure.Patient was stable while in the lab.Unknown if patient required extended hospitalization because of the adverse event.Relevant tests/laboratory data-echo.Other relevant history-unknown.Generator information-unknown.Transseptal puncture was performed with a baylis needle.No evidence of steam pop.The event occurred during the transseptal phase.No radio frequency was given.The device is not available for return.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF BI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez
MX  
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16076207
MDR Text Key306408075
Report Number2029046-2022-03258
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberD134805
Device Catalogue NumberD134805
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NON BWI ¿ AGILIS SHEATH.; NON BWI ¿ BAYLIS NEEDLE.; UNKNOWN BRAND GENERATOR.
Patient Outcome(s) Required Intervention; Life Threatening;
Patient SexFemale
-
-