MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-02H
Device Problems Mechanical Problem (1384); Noise, Audible (3273); Intermittent Communication Failure (4038)
Patient Problems Insufficient Information (4580); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 12/06/2022
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing no sound.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed sliced silicone overmold on the top and bottom covers, as well as a severed electrode.These anomalies are believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was obtained intermittently.The condition of the electrode prevented and electrical test from being performed.The device passed the electrical test performed.The residual gas analysis test exceeded the testing limit.This device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issues from one feedthru vendor.An internal corrective action was implemented.Feedthru assemblies from this vendor are no longer used.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
heidi white
28515 westinghouse place
valencia, CA 91355
MDR Report Key16076305
MDR Text Key306504573
Report Number3006556115-2022-02169
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 12/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/01/2007
Device Model NumberCI-1400-02H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2022
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-