Model Number FINETUNER ECHO [FINETUNER 2] |
Device Problems
No Device Output (1435); Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Event Description
|
The finetuner echo was received at service and repair and the repair request form (rrf) stated that there was a mechanical problem.The user reported that it was difficult to remove the battery and the screen was blocked.
|
|
Event Description
|
The finetuner echo was received at service and repair and the repair request form (rrf) stated that there was a mechanical problem.The user reported that it was difficult to remove the battery and the screen was blocked.
|
|
Manufacturer Narrative
|
Conclusion: according to the information received, a finetuner echo was sent to service repair due to not functioning.During device investigation a failed capacitor was detected which was clearly the reason for the reported issue.Electrical inspection could not be performed because of the observed malfunction.Also, other components were missing from the circuit board.This is a final report.
|
|