Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Loss of consciousness (2418); Convulsion/Seizure (4406)
|
Event Date 12/25/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the adc device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer received treatment of juice, sugar, or food provided by a non-healthcare professional.Attempts to contact the customer multiple times up to this point has been unsuccessful.No further information was provided.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Dhrs for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the adc device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer received treatment of juice, sugar, or food provided by a non-healthcare professional.Attempts to contact the customer multiple times up to this point has been unsuccessful.No further information was provided.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|