Catalog Number 329489 |
Device Problem
Complete Blockage (1094)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/18/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that the bd micro-fine¿ + pen needle was blocked during use.The following information was provided by the initial reporter, translated from chinese: "during use, it is found that the needle is blocked and cannot be used normally".
|
|
Event Description
|
It was reported that the bd micro-fine¿ + pen needle was blocked during use.The following information was provided by the initial reporter, translated from chinese: "during use, it is found that the needle is blocked and cannot be used normally".
|
|
Manufacturer Narrative
|
H6: investigation summary: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.Dhr was reviewed and there were not any qns or other events that could be related to this complaint.Clog test was conducted on 7pcs retention samples and all no needle clog fail found.Base on investigation above, the certain cause cannot be concluded at the moment.
|
|