Model Number LSP112V
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Event Description
It was reported the asymptomatic patient presented for a leadless pacemaker implant.During post-operation check, the ventricular leadless pacemaker had no conductive telemetry.Troubleshooting was attempted, and communication was eventually successful, however the physician elected to replace the device.The patient was stable.
 
Event Description
It was reported the patient presented for a leadless pacemaker implant.During post-operation check while still in the operating room, the ventricular leadless pacemaker had no conductive telemetry.Trouble shooting was attempted, and communication was eventually successful, however the physician elected to replace the device.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVEIR¿ LEADLESS PACEMAKER, RIGHT VENTRICLE
Type of Device
LEADLESS PACEMAKER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16085672
MDR Text Key306502491
Report Number2017865-2023-00082
Device Sequence Number1
Product Code PNJ
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P150035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/25/2023
Device Model NumberLSP112V
Device Catalogue NumberLSP112V
Device Lot NumberS000087722
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age69 YR
Patient SexMale
-
-