Model Number N/A |
Device Problems
Positioning Failure (1158); Difficult to Insert (1316)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the liner would not engage in the cup upon impaction.Another liner was opened, and the impaction was successful.No debris fell nor was there any consequences or impact to the patient.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were {update/corrected} updated: d9; g3; h2; h3; h6 visual examination of the returned product identified scratches across the outer radius.The side wall is scuffed and scratched.The inner radius exhibits nicks and scratches.It is unknown if the damage occurred prior to or during assembly attempts the additional information does not change the outcome of the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further information at the time of this report.
|
|
Search Alerts/Recalls
|