Model Number A22040A
Device Problems Break (1069); Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
Olympus was informed that the ceramic beak had a dent during reprocessing.No further information was provided but there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
No investigation will be performed on the returned device as this has been determined a duplicate complaint.The device evaluation can be found on mfr number 9610773-2023-00051.The evaluation at omsi confirmed the ceramic insulation at the distal end of the inner sheath to be damaged.The type of damage found is typically caused by thermo-mechanical fatigue caused by wear and tear and/or mechanical overload, possibly as a result of excessive force.Therefore, this event/incident was attributed to use error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH
Type of Device
HYSTEROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16100406
MDR Text Key308706993
Report Number9610773-2023-00091
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761029339
UDI-Public04042761029339
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040A
Device Catalogue NumberA22040A
Device Lot Number186W -0383
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2022
Was the Report Sent to FDA? No
Date Manufacturer Received12/06/2022
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-