Analysis and results: there are no previous complaints of this code batch of which we manufactured and distributed in the market 2,232 units of this code batch.There are no units in stock in b.Braun surgical's warehouse.Without closed and/or defective samples a proper analysis cannot be performed.Reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Remarks: as indicated in the instructions for use of the product: "when working with optilene® suture materials great care should be taken to ensure that the use of surgical instruments, such as forceps and needle holders, do not cause any crushing or crimping damage to the suture material".Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Actions on product distributed of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|