MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN G7 LINER; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN G7 LINER; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Naturally Worn (2988)
Patient Problem Pain (1994)
Event Date 12/15/2022
Event Type  Injury  
Event Description
It was reported that a right hip revision was performed due to liner wear and pain during high activity movement.The procedure was completed using another g7 liner and ceramic head.There is no additional information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2023-00020.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, as supplemental medwatch will be submitted.
 
Event Description
There is no update to the original complaint description provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the provided pictures identified the liner is heavily damaged from removal.Based on image liner wear cannot not be decerned from the removal damage and debris present.The femoral head have what appear to be wear lines, scuffing and bio-debris present.No additional detail can be seen based on provided images.Part and lot identification are necessary for review of device history records, neither were provided.Radiographs were provided and reviewed by a health care professional.The image was reviewed and determined as it is a single undated image that would not enhance the investigation.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN G7 LINER
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16104625
MDR Text Key306730264
Report Number0001825034-2023-00019
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberUNK G7 LINER
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/17/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
UNK HEAD
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexMale
Patient Weight154 KG
-
-