3 of 3 reports.Other mfg report numbers: 3014334038-2022-00293, 3013886523-2022-00593.A facility reported a microsensor was implanted on (b)(6) 2022 in a male patient of approximate 8 years-old.A few minutes after implantation the icp readings jumped from clinical to 40.Due to the issue the microsensor was replaced.The patient went to picu (pediatric intensive care unit) with normal icp reading, however, approximately 1 hour later the sensor jumped to 40; then the second microsensor was explanted.It was also found that icp express cable was faulty (id: (b)(4) but not the monitor.
|