Catalog Number 626631
Device Problem High Readings (2459)
Patient Problem Insufficient Information (4580)
Event Date 12/15/2022
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
2 of 3 reports.Other mfg report numbers: 3014334038-2022-00293, 3013886523-2023-00001.A facility reported a microsensor was implanted on (b)(6) 2022 in a male patient of approximate 8 years-old.A few minutes after implantation the icp readings jumped from clinical to 40.Due to the issue the microsensor was replaced.The patient went to picu (pediatric intensive care unit) with normal icp reading, however, approximately 1 hour later the sensor jumped to 40; then the second microsensor was explanted.It was also found that icp express cable was faulty (id: (b)(4) but not the monitor.
 
Manufacturer Narrative
The microsensor was not returned for evaluation after three good faith attempts (gfes) were made.Lot number information has been provided; therefore, device history record (dhr) was reviewed, and no anomalies were found.The root cause of the reported issue could not be determined.However, a probable root cause for the reported complaint is due to excessive pressure applied to product.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROSENSOR BASIC KIT R
Type of Device
ICP MICROSENSORS (ROHS)
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16107030
MDR Text Key306746316
Report Number3013886523-2022-00593
Device Sequence Number1
Product Code GWM
Combination Product (y/n)N
PMA/PMN Number
K153347
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2023
Device Catalogue Number626631
Device Lot Number5967909
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/24/2023
Date Device Manufactured11/15/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexMale
-
-