Model Number N/A |
Device Problems
Detachment of Device or Device Component (2907); Intermittent Loss of Power (4016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).(b)(6).Evaluation and investigation is in process.Once the investigation is complete, a supplemental report will be filed.
|
|
Event Description
|
It was reported that there was a loss of power.The hand piece is working but it releases the reciprocating saw attachment without pushing button pressure.There was no harm or delay reported.Due diligence is in progress to date no further information has become available.No adverse events were reported as a result of this malfunction.
|
|
Event Description
|
Additional information reported that there was a 20 minute delay and another device was used to complete the procedure.
|
|
Manufacturer Narrative
|
This complaint is recorded by zimmer biomet under (b)(4).If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no additional event information available.
|
|
Manufacturer Narrative
|
G2: foreign country: italy.Review of the most recent repair record determined the motor was seized.The motor was replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|