Model Number 9-PFO-2518
Device Problem Malposition of Device (2616)
Patient Problem Pericardial Effusion (3271)
Event Date 12/14/2022
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2022, a 25/18mm amplatzer talisman pfo occluder was selected for implantation.Heparin was administered.During the procedure, the device was prematurely deployed in the left atrium, but not released from the delivery system.The device then damaged the heart with the left disc, and then was retrieved using the delivery system and removed from patient anatomy.The patient was diagnosed with a pericardial effusion around the left atrium.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient is reported to be stable.No additional information was provided.
 
Manufacturer Narrative
An event of the device was deployed prematurely during procedure and pericardial effusion around the left atrium was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TALISMAN PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16118059
MDR Text Key306878550
Report Number2135147-2023-00093
Device Sequence Number1
Product Code MLV
UDI-Device Identifier05415067033314
UDI-Public05415067033314
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/06/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9-PFO-2518
Device Catalogue Number9-PFO-2518
Device Lot Number8566716
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-