• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751586
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported the trocar was stuck in the vitrector during a vitrectomy surgery.When it was tried to take out it uprooted the trocar out from the site.The vitrector and the trocar was replaced and the surgery as completed.There was no report of a patient harm.
 
Manufacturer Narrative
The suspect product has been updated.The manufacturer internal reference number is: (b)(4).
 
Event Description
The suspect product has been updated.
 
Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation; therefore, the condition of the product could not be verified.No lot number was identified with this complaint; therefore, a device history record review could not be conducted.A sample was not received at the manufacturing site and no lot information was provided; therefore, the root cause for the customer complaint issue cannot be determined.The cause of the reported event cannot be determined with the information obtained; therefore, specific action with regards to this complaint cannot be taken.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16131182
MDR Text Key308754673
Report Number2028159-2023-00032
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Remedial Action Notification
Type of Report Initial,Followup,Followup
Report Date 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065751586
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/04/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CUSTOM-PAK SURGICAL PROCEDURE PACK.; SURGICAL PRODUCT, UNSPECIFIED.; SURGICAL PRODUCT, UNSPECIFIED.
-
-