• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ACCU-CHEK INFORM II TEST STRIPS; BLOOD GLUCOSE MONITORING TEST STRIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ACCU-CHEK INFORM II TEST STRIPS; BLOOD GLUCOSE MONITORING TEST STRIPS Back to Search Results
Catalog Number 05942861001
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/15/2022
Event Type  malfunction  
Manufacturer Narrative
Qc passed on the meter within 24 hours of the event.The product was requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow-up report will be submitted.On a regular basis, inform strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.
 
Event Description
The initial reporter complained of discrepant glucose results for 2 patients tested with accu-chek inform ii meter serial number (b)(4)."patient 2" was tested at 12:28 (capillary fingerstick) with a result of 64 mg/dl.The patient was re-tested at 12:35 (capillary fingerstick) with a result of 100 mg/dl."patient 1" was tested at 12:31 (capillary fingerstick) with a result of 260 mg/dl.The patient was re-tested at 12:41 (capillary fingerstick) with a result of 161 mg/dl.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-CHEK INFORM II TEST STRIPS
Type of Device
BLOOD GLUCOSE MONITORING TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16132410
MDR Text Key308727018
Report Number1823260-2023-00097
Device Sequence Number1
Product Code LFR
UDI-Device Identifier00365702428102
UDI-Public00365702428102
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121679
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05942861001
Device Lot Number670208
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/15/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-