Model Number AIA-360 IMMUNOASSAY ANALYZER
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2022
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and was not able to duplicate the problem.While troubleshooting, fse attempted to look for something internal that could have restricted the turn table movement and noticed the flag for the turntable sensor was close to hitting the sensor.Fse resolved the complaint by adjusting the turntable sensor flag to ensure the flag was positioned properly in the sensor.Fse repaired and validated the analyzer by successfully performing quality control run without error and within acceptable range.No further action required by field service.The aia-360 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 21oct2021 through aware date 21nov2022.There were no similar complaints identified during the search period.The aia-360 operator's manual under section 7-1: error messages states the following: (4002) turn table home sensor description: the home position of the turntable motor cannot be detected.Troubleshooting: turn the power off and on again.If this problem reoccurs, contact the service department.(4004) turn table home slip description: the turntable motor slipped.Troubleshooting: turn the power off and on again.If this problem reoccurs, contact the service department.The most probable cause of the reported event was due misaligned turntable sensor.
 
Event Description
A customer reported error message ¿4002 turn table home sensor and 4004 turn table home slip¿ on the aia-360 analyzer.Technical support specialist (tss) instructed the customer to turn the analyzer off and on, but the error persisted and a loud grinding noise occurred.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for estradiol (e2).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key16135140
MDR Text Key308838243
Report Number3004529019-2023-00231
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189282896
UDI-Public04560189282896
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360 IMMUNOASSAY ANALYZER
Device Catalogue Number019945
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-