Catalog Number 8606500 |
Device Problems
Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 12/14/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still on-going.The results will be provided with a follow-up report.
|
|
Event Description
|
It was reported that the device autonomously switched from pressure controlled ventilation to manual mode during a case.No patient injury reported.
|
|
Manufacturer Narrative
|
Log file evaluation performed by the manufacturer confirms the reported behavior: the supervisor function detected a sporadic loss of the signal from the sensor that monitors the inspiratory cylinder pressure.To protect the patient from potentially hazardous output the device is designed to shut-down automatic ventilation and to post a corresponding "ventilator failure" alarm to alert the user.Manual ventilation with the built-in breathing bag remains possible; the monitoring functions are still available as well.As confirmed for the particular case automatic ventilation could be continued two minutes later.The failure only occurred once without being able to determine the exact root cause for the temporary signal loss of the pressure sensor.As a preventive measure the ventilator assembly and the pcb for ventilation and gas control (including the affected pressure sensor as it is onboard) were replaced.The unit was successfully tested afterwards and was returned to use without further problems reported.Dräger finally concludes that the workstation responded as designed upon the deviation.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|
|
Event Description
|
It was reported that the device autonomously switched from pressure controlled ventilation to manual mode during a case.No patient injury reported.
|
|
Search Alerts/Recalls
|