• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-344-10-K
Device Problems Signal Artifact/Noise (1036); Impedance Problem (2950)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Pending investigation of returned lead.
 
Event Description
The patient's right depth lead displayed abnormal signals, saturation, and flat line activity during review of patient ecog data.The abnormal activity appears to have begun around (b)(6) 2022.These activities are indicative of a right depth lead break.The lead was explanted and replaced on (b)(6) 2022.During explant the surgeon reported evidence of lead damage near the burr hole cover.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key16138373
MDR Text Key308719142
Report Number3004426659-2022-00054
Device Sequence Number1
Product Code PFN
UDI-Device Identifier00855547005298
UDI-Public010085554700529817241014
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberDL-344-10-K
Device Catalogue Number1007931
Device Lot Number31981-1-1-1
Date Manufacturer Received12/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age35 YR
Patient SexMale
-
-