MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number D-EVOLUTFX-34
Device Problems Material Separation (1562); Failure to Advance (2524)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2023
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during implant of this transcatheter bioprosthetic valve via the right femoral artery that was diseased and tortuous, the delivery catheter system (dcs) was unable to advance beyond the common iliac artery due to calcium.The dcs was rotated and reattempted to advance, however was unsuccessful.It was determined to pull the dcs out however when retracting the system, the nose cone of the dcs broke off remaining in the calcium and the proximal portion separated.A dilator was used to advance the nose cone into the descending aorta.A 22 french sheath was inserted, followed by a snare through the sheath.The nose cone was snared into the sheath and both devices were pulled out of the patient at once.A new 22 french sheath was introduced.A new valve and dcs were used for a successful implant.No damage occurred to the iliofemoral artery.No adverse patient effects were reported.
 
Manufacturer Narrative
Product analysis: upon receipt at medtronic's quality laboratory, the delivery catheter system (dcs) was received with a valve loaded within the capsule.The handle appeared intact.The nosecone was received detached from the inner member shaft.On retraction of the capsule via the rotation of the deployment knob, the valve deployed without issue.The deployment knob appeared to retract and advance the capsule.The trigger moved to fully advanced and retracted positions and locked in place when released.The tip-retrieval mechanism appeared intact.The device was returned with the end cap/screw gear snap fit connected.The capsule appeared intact with no evidence of damage.The distal detachment site had part of the inner member shaft remaining within the nosecone and had fraying to the inner member threading.The proximal detachment site had frayed inner member threading.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Updated data: h6.Eval code method eval code result - the fda.Gov website shows a cause investigation - investigation findings code of c24 (malfunction was verified but no conclusive finding is available); however, this code is not available for selection on this electronic regulatory report.Code c22 was selected, but c24 is an accurate code to apply for this event.Eval code conclusion conclusion: the complaint device was received for analysis and the device separation was confirmed.A device history record review was performed on the device and there were no correlations / issues identified regarding manufacturing.The device was manufactured per approved and released manufacturing processes and the device met all applicable manufacturing specifications prior to release for distribution.The reported event indicates that the nose cone of the dcs broke off remaining in the calcium and the proximal portion separated.It is possible that there was a false cut present on the inner member and coupled with the interaction with the calcified and tortuous anatomy subsequently resulted in the nose cone detaching, however, this cannot be conclusively confirmed.An assessment by the vendor was performed and it was noted that the issue may have occurred due to a false cut.A false cut occurs when the blade does not cut through the entire wall of the tubing, and this could occur during the trimming operation at microlumen; however, this cannot be conclusively confirmed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX DCS
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16147285
MDR Text Key307184645
Report Number2025587-2023-00104
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 09/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/12/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD-EVOLUTFX-34
Device Catalogue NumberD-EVOLUTFX-34
Device Lot Number0011454408
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received09/19/2023
Date Device Manufactured10/14/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-