• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI CAPSURE Z NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI CAPSURE Z NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5054-58
Device Problems Break (1069); Device Contamination with Body Fluid (2317)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the right ventricular (rv) lead exhibited a breach on the outer insulation of the pacing lead due to depression with blood ingression noted.The rv lead was explanted when the patient passed away with their death noted to be unrelated to performance of the rv lead.
 
Manufacturer Narrative
Concomitant medical products: addr01 ipg implanted: (b)(6) 2016.Product event summary: the proximal portion of the lead was returned and analyzed.The outer insulation of the lead developed a breach due to a depression while in vivo.The outer insulation of the lead was observed to have blood ingression.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE Z NOVUS LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16156236
MDR Text Key308449555
Report Number2649622-2023-01111
Device Sequence Number1
Product Code DTB
UDI-Device Identifier00681490123228
UDI-Public00681490123228
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P850089
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 01/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/13/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/05/2009
Device Model Number5054-58
Device Catalogue Number5054-58
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2022
Date Manufacturer Received01/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/10/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
555453 LEAD
Patient SexMale
-
-