Initial reporter occupation is patient/consumer.The meter was returned for investigation.The test strips were not returned.A full display check was performed and revealed no missing or faded segments.No display issues were observed.The returned meter was tested using retention strips and retention controls.Testing results (qc range = 4.1 - 6.8 inr): qc 1: 5.1 inr, qc 2: 5.2 inr, qc 3: 5.0 inr.The obtained qc values were in the allowed range of the used combination strip lot - qc lot.All measurements were without error messages.The patient alleged receiving error 5.A review of the meter error log revealed error 5.Error 5 occurs when the applied sample volume is insufficient or when sample detection is uncertain.Generally, this is caused by wrong handling, or by meter contamination.Product labeling for error 5 states: "error: blood application error applying blood to the test strip or blood sample size was too small.Solution: turn the meter off and remove the test strip.Re-read the instructions about applying blood (starting on page 33).Repeat the test with a new test strip and a sufficient amount of blood from a different finger.If you still get the error message, call roche customer support center at (b)(4).".A review of the meter error log revealed no result of 5.0 inr.The patient had mistaken an error 5 for a 5.0 inr result.Per product labeling, "coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods." the investigation did not identify a product problem.The cause of the event could not be determined.
|