Catalog Number 2C2063K |
Device Problems
Fluid/Blood Leak (1250); Material Split, Cut or Torn (4008)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the fill port of a large volume infusor was damaged and resulted in a fluid leak.This issue was further described as, ¿they noticed there was a leak at the connection point it looks like there is a split in the hard plastic connector¿.This event occurred while compounding the device.The device was filled with sodium chloride 180ml.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
Initial reporter postal code: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Correction: g4: pma/510k #.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.Visual inspection was performed to the photograph which showed white drug residue around the fill port area which suggests a leak may have occurred.Additionally, the photo showed cracks on the fill port which resulted in leak.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|