SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTILATORS PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
|
Back to Search Results |
|
Model Number P310NUS |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the device is leaking.There has been no report of patient involvement, no observable clinical symptoms, or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
Operator of device is unknown.No information has been provided to date.A product sample was received and is awaiting evaluation.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Manufacturer Narrative
|
One device was returned for evaluation with a damaged front panel, all small knobs were cracked, housing case had minor to major cracks and manometer is out of spec.A lock/off leak test was performed and a leak was detected from the demand valve.Confirming the reported complaint.A faulty demand valve was the root cause of the issue.Once replacing the demand valve the device passed the functional tests.A manufacturing dhr is not relevant due to the age of the device and impact damage sustained.
|
|