ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP SCISSORS; SCISSORS, OPHTHALMIC Back to Search Results
Catalog Number 705.52P
Device Problem Failure to Cut (2587)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/09/2022
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported that before a vitrectomy an ophthalmic scissor was not sharp and it failed to cut.Surgery was completed with another product.There was no harm to the patient.
 
Manufacturer Narrative
A review of the device history record (dhr) traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The sample is received with inner blister, outer blister and cover foil showing the lot information.The sample shows no macroscopic signs of damage.The sample shows signs of surgery residues and is returned in a closed status.The sample was visually inspected with the aid of a photomicroscope with various magnifications.The sample was functionally tested.It was noticed that the device gets stuck in closed position, hence on this it was not able to cut.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.It cannot be determined how or where the damage to the sample occurred; therefore a root cause for the customers reported event could not be determined.However, the damage was consistent with damage that can occur due to improper handling.The exact root cause for the customers reported event is unknown, therefore, specific action cannot be taken.Complaints are reviewed and monitored at regular intervals for adverse trends.No adverse trends have been observed associated with the reported product and event.No action has been identified for this reported event.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP SCISSORS
Type of Device
SCISSORS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16194571
MDR Text Key307581078
Report Number3003398873-2023-00010
Device Sequence Number1
Product Code HNF
UDI-Device Identifier07612717070691
UDI-Public07612717070691
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Catalogue Number705.52P
Device Lot NumberF197850
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-