Model Number 75175-83 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Blurred Vision (2137); Dizziness (2194); Shaking/Tremors (2515)
|
Event Date 12/20/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The available tripped trend reports were reviewed for the reported complaint and fs libre reader and no trends were identified that would indicate any product related issues.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A caregiver reported an "error-7" message with the adc device.Due to this issue, customer was unable to monitor glucose and experienced symptoms of hypoglycemia including"loss of vision, dizziness, tremors" and was provided treatment with three spoons of sugar by a non-healthcare professional.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Repeated attempts by adc to retrieve the product were unsuccessful and/or the customer discarded the product.At this time product has not yet been returned and a valid serial number has not been provided.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The available tripped trend reports were reviewed for the reported complaint and freestyle optium neo meter and no trends were identified that would indicate any product related issues.The device mfg date is unknown.The date entered in section h4 is the date abbott diabetes care became aware of the event.If the product is returned, a physical investigation will be performed and a follow-up report submitted.This serves as a correction report.Section h10 (addtl mfg narrative) was incorrectly documented in the initial report.The correction has been made here.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
A caregiver reported an "error-7" message with the adc device.Due to this issue, customer was unable to monitor glucose and experienced symptoms of hypoglycemia including"loss of vision, dizziness, tremors" and was provided treatment with three spoons of sugar by a non-healthcare professional.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|