• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALUNG TECHNOLOGIES, INC. HEMOLUNG RAS; EXTRACORPOREAL SYSTEM FOR CARBON DIOXIDE REMOVAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALUNG TECHNOLOGIES, INC. HEMOLUNG RAS; EXTRACORPOREAL SYSTEM FOR CARBON DIOXIDE REMOVAL Back to Search Results
Model Number HL-CR4-01-000
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Anemia (1706)
Event Date 07/16/2019
Event Type  Injury  
Manufacturer Narrative
Patient information was not provided.Alung technologies, inc.Manufactures the hemolung ras and catheter.The incident occurred in albany, new york.Medical intervention was required to prevent further patient injury.An adverse event was reported in which the subject experienced episodic anemia.The data log from hemolung therapy has been retrieved and reviewed by both clinical and software engineering staff.During hemolung therapy, a 102 error occurred in which the o2 sensor value was greater than 24%.The controller software operated as intended by stopping the pump.Technical support was provided to the clinical staff in which they power cycled device and chose to continue the hemolung therapy at their discretion.No abnormalities were noted within the co2 removal and blood flow graphs.Hemolung therapy was provided as intended.No capa was opened because of this eveny.Episodic anemia is a known possible occurrence with extracorporeal hemolung therapy.Review of the controller data log shows that hemolung therapy was provided as intended.Alung technologies, inc.Will continue to monitor any issues as they are reported.This mdr is being filed in response to a retrospective view of all alung technologies, inc.Complaints, which identified that the reporting decision for this complaint was not correct.This mdr is being filed retrospectively to correct the error in the reporting decision.
 
Event Description
Alung technologies, inc.Receieved a report of a patient experiencing anemia during hemolung therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOLUNG RAS
Type of Device
EXTRACORPOREAL SYSTEM FOR CARBON DIOXIDE REMOVAL
Manufacturer (Section D)
ALUNG TECHNOLOGIES, INC.
2500 jane st
suite 1
pittsburgh PA 15203
Manufacturer Contact
ryan coyle
2500 jane st
suite 1
pittsburgh, PA 15203
MDR Report Key16200012
MDR Text Key307658615
Report Number3009763347-2023-00046
Device Sequence Number1
Product Code QOH
UDI-Device Identifier00850046004346
UDI-Public(01)00850046004346(17)171001
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN210006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberHL-CR4-01-000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/16/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-