Alung technologies, inc.Manufactures the hemolung ras and catheter.The incident occurred in (b)(4), united kingdom.Shortly following initiation of hemolung therapy, the controller alarmed with elevated cabinet o2.O2 was running as sweep gas.The staff was advised to switch to using air for the sweep gas, which resolved the issue.However, four hours later hemolung therapy was attempted to return to o2 for the sweep gas as the patient was hypoxic.But the controller began again alarming with elevated cabinet o2.Support was provided in which the staff was advised to continue hemolung therapy with room air as the sweep gas.The data log was retrieved and reviewed by both clinical and software engineering staff.The software engineering review suggest a possible leak in the vacuum pump / sweep gas module, as opposed to a faulty o2 sensor.The controller was exchanged and repaired.The clinical review shows no abnormalities were noted within the co2 removal and blood flow graphs.Hemolung therapy was provided as intended.No capa was opened because of this event.Alung technologies, inc.Will continue to monitor any issues as they are reported.This mdr is being filed in response to a retrospective view of all alung technologies, inc.Complaints, which identified that the reporting decision for this complaint was not correct.This mdr is being filed retrospectively to correct the error in the reporting decision.
|