• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number IBI-81104
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2023
Event Type  malfunction  
Event Description
During a supraventricular tachycardia procedure, the physician had difficulty trying to access the cs with the catheter resulting in a delay of the procedure.The catheter was noticed to be deformed (s shape) when the catheter was taken out of the body for checking.The catheter was replaced with another ibi-81104 and had the same issue.The catheters were then replaced with a livewire catheter and the issue was resolved.The case proceeded as per normal without any patient consequences.
 
Manufacturer Narrative
One decapolar, inquiry steerable diagnostic catheter was received for evaluation.The catheter created an ¿s¿ shaped curve during deflection, which no longer matched the required curve template.The catheter met specifications for electrical testing.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The root cause for an ¿s¿ shaped curve is due to the activation wire shifting over the flat wire.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INQUIRY¿ STEERABLE DIAGNOSTIC CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
2375 morse ave
irvine CA 92614
Manufacturer (Section G)
ST. JUDE MEDICAL
2375 morse ave
irvine CA 92614
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16200667
MDR Text Key308469681
Report Number2030404-2023-00004
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05414734302975
UDI-Public05414734302975
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K961924
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIBI-81104
Device Catalogue NumberIBI-81104
Device Lot Number8521654
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-