• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH, WITH A22085A; HYSTEROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH, WITH A22085A; HYSTEROSCOPE Back to Search Results
Model Number A22040T
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/19/2022
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.
 
Event Description
It was reported to olympus that the beak of the resectoscope fell into the bladder during the turis operation.It was retrieved with foreign body forceps.There was no harm or user injury reported due to the event.
 
Manufacturer Narrative
This medwatch is being supplemented with additional information obtained and the manufacturer's final investigation results.During investigation, it was confirmed that the ceramic beak of the inner sheath has broken out.Based on the present damage pattern, we assume that the insulation tip's fracture was caused by thermal and mechanical induced impact.A manufacturing and quality control review was performed for the affected serial number without showing any non-conformities or deviations regarding the described issues.The ifu carries a warning that the ceramic tip can break due to mechanical loading or thermally induced straining.Thus, it is the responsibility of the user to inspect the instrument prior to every procedure: the instructions for use state: ¿4 before use: warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: no corrosion; no dents; no scratches.Ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿ olympus will continue to monitor the field performance of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INNER SHEATH, FOR 26 FR. OUTER SHEATH, WITH A22085A
Type of Device
HYSTEROSCOPE
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16210045
MDR Text Key307798203
Report Number9610773-2023-00276
Device Sequence Number1
Product Code HIH
UDI-Device Identifier04042761029360
UDI-Public04042761029360
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K931994
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA22040T
Device Catalogue NumberA22040T
Device Lot Number19Y07
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-