OLYMPUS WINTER & IBE GMBH INNER SHEATH, FOR 26 FR. OUTER SHEATH, WITH A22085A; HYSTEROSCOPE
|
Back to Search Results |
|
Model Number A22040T |
Device Problem
Device Dislodged or Dislocated (2923)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/19/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not returned to olympus.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.
|
|
Event Description
|
It was reported to olympus that the beak of the resectoscope fell into the bladder during the turis operation.It was retrieved with foreign body forceps.There was no harm or user injury reported due to the event.
|
|
Manufacturer Narrative
|
This medwatch is being supplemented with additional information obtained and the manufacturer's final investigation results.During investigation, it was confirmed that the ceramic beak of the inner sheath has broken out.Based on the present damage pattern, we assume that the insulation tip's fracture was caused by thermal and mechanical induced impact.A manufacturing and quality control review was performed for the affected serial number without showing any non-conformities or deviations regarding the described issues.The ifu carries a warning that the ceramic tip can break due to mechanical loading or thermally induced straining.Thus, it is the responsibility of the user to inspect the instrument prior to every procedure: the instructions for use state: ¿4 before use: warning infection control risk properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing inspecting the product visually inspect the product.Make sure that it has: no corrosion; no dents; no scratches.Ceramic insulation at distal end visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning risk of injury impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿ olympus will continue to monitor the field performance of this device.
|
|
Search Alerts/Recalls
|
|
|