Model Number CDI510H |
Device Problem
Failure to Recalibrate (1517)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/27/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Terumo has not received the device for evaluation; therefore, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.Component code: 510 - sensor health effect ¿ impact code: 2199 - no health consequences or impact.Health effect ¿ clinical code: 4582 - no clinical signs, symptoms or conditions.Medical device problem code: 1517 - failure to recalibrate.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
|
|
Event Description
|
The user facility reported to terumo cardiovascular that during cardiopulmonary bypass, there was a recalibration error.About 10 times of recalibration errors occurred during use with cdi550.The message shown was chosen blood gas module configuration does not match placement of cable heads in calibrator.A backup shunt sensor was used to continue the procedure.It is unknown if there was a blood loss during this case, and that terumo is attempting to gather the missing information.*no consequence or health impact to patient *product was changed out *procedure was completed successfully.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations.Upon further investigation of the reported event, the following information is new and/or changed: h6 (identification of evaluation codes 10, 11, 3331, 213, 67).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #3: 3331 - analysis of production records.Investigation findings: 213 - no device problem found.Investigation conclusions: 67 - no problem detected.The returned sample was visually inspected upon receipt with no anomalies noted.A new buffer solution was added to the returned sample and it was then attempted for gas calibration using the cdi 500 and cdi540.It was able to be gas calibrated successfully with no issues.A retention sample from the same product code and lot number was found to be successfully gas calibrated with no issues.The complaint and retention sample were then circulated with buffer solution and found to produce consistent values on the cdi500 monitor.The samples were found to perform as expected during the investigation.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|
Manufacturer Narrative
|
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on january 20,2023.Upon further investigation of the reported event, the following information is new and/or changed: h3 (device evaluation anticipated by manufacturer - a second follow-up will be submitted upon completion of the investigation and/or submission of new information, thus tcvs references conclusions code 11.) all available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|
|