Model Number P7900B100010
Device Problem Smoking (1585)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the bed usb port was smoking.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hillrom technician found the usb port needed to be replaced.The external alarm is the source of the brake not set alarm and local bed exit alarm.Per the hillrom service manual, do not allow the conductive parts of the patient charging cable, to come into contact with other conductive parts, this includes earth ground.Otherwise, an electrical short might occur, risking electric shock to the patient and damage to their portable electronic device.The usb charging port may not be compatible with all devices.The usb charging port can be used to charge personal electronic devices (ped) such as cell phones, computer tablets, portable media players, and pocket computers.The port is not designed for devices that require less than 170 ma of power such as usb reading lights,but supports up to 2.4 a of current.A search of the hillrom maintenance records did not show hillrom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the usb port to resolve the reported event.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRELLA MED-SURG
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
kayla miller
1069 state route 46 east
batesville, IN 47006
8129310130
MDR Report Key16223431
MDR Text Key308477937
Report Number1824206-2023-00040
Device Sequence Number1
Product Code FNL
UDI-Device Identifier00887761985162
UDI-Public010088776198516211200124
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 01/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP7900B100010
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-