Model Number LUCAS
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/14/2021
Event Type  Death  
Manufacturer Narrative
This mdr is being submitted to correct the incorrect manufacturing site and related fields that were previously submitted under 3005445717-2021-00010.Due to character limitations initial reporter phone, was left blank.The initial reporter¿s phone number is (b)(6).Physio-control contacted the customer to request additional information on the patient.The customer provided physio-control with the available patient information.Patient fields in which information is not provided were intentionally left blank.   though physio-control requested some patient information, physio will not request any information which can identify, directly or indirectly, a person to be in accordance with regulation (eu) 2016/679 of the european parliament and of the council.A physio-control service representative performed an initial evaluation of the customer¿s device and was not able to verify the reported issue.After completing some unrelated repairs, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer.Clinical review of the reported event was performed.It can not be excluded that the failing of the device contributed to the patient death even if the survival rate in this transported patient population is low.
 
Event Description
A customer contacted physio-control to report that their device failed during intervention on a patient.The device alarmed and stopped.The patient did not survive reported event.The healthcare provider confirmed that the device use contributed to the patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCAS 2, 2.1 SW, IN SHIPPING BOX,EN,EN INTL
Type of Device
COMPRESSOR, CARDIAC, EXTERNAL
Manufacturer (Section D)
JOLIFE AB - 3005445717
scheelevagen 17
ideon science park
lund SE-22 3 70
SW  SE-223 70
Manufacturer (Section G)
JOLIFE AB - 3005445717
scheelevagen 17
ideon science park
lund SE-22 3 70
SW   SE-223 70
Manufacturer Contact
todd bandy
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key16237076
MDR Text Key308063582
Report Number3005445717-2023-00031
Device Sequence Number1
Product Code DRM
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K090422
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/24/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberLUCAS
Device Catalogue Number99576-000025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/16/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age37 YR
Patient SexMale
-
-