This event has been recorded by zimmer biomet under (b)(4).Investigation is complete.This is an initial final report submission.Review of the most recent repair record determined the motor speed was unstable and the switch, screws, compression cap, and end cap were damaged.The motor, screws, switch, compression cap, and end cap were replaced and resolved the reported issue.Review of the previous repair record identified no related repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|