Model Number N/A
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Investigation is complete.This is an initial final report submission.Review of the most recent repair record determined the motor speed was unstable and the switch, screws, compression cap, and end cap were damaged.The motor, screws, switch, compression cap, and end cap were replaced and resolved the reported issue.Review of the previous repair record identified no related repairs to the reported event.The device was tested and found to be functioning to specifications prior to release to the customer.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that the device does not function well.No adverse events were reported as a result of this malfunction.Due diligence is complete and no additional information is available.At product evaluation investigation the motor speed was unstable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDPIECE, ELECTRIC DERMATOME
Type of Device
DERMATOME LINE POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16240073
MDR Text Key308645407
Report Number0001526350-2023-00089
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00882100100
Device Lot Number60615584
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/30/2022
Was the Report Sent to FDA? No
Date Manufacturer Received01/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
-
-