• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Defective Component (2292)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.This report is being submitted as the result of a retrospective review conducted in capa 584165.
 
Event Description
It was reported that during schedule service, one of the led charge indicators for the battery (the 4th one from the bottom up) in cardiosave intra-aortic balloon pump (iabp) was defective.There was no patient involvement.
 
Manufacturer Narrative
A getinge field service engineer (fse) that encountered the issue, replaced the batteries.Unit passed all functional and safety test per factory specifications.The fse performed a calibration to the unit and passed all functional and safety tests per factory specifications.Iabp was returned to customer and cleared for clinical use.
 
Event Description
It was reported that during schedule service, one of the led charge indicators for the battery (the 4th one from the bottom up) in cardiosave intra-aortic balloon pump (iabp) was defective.Patient involved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16241314
MDR Text Key308203708
Report Number2249723-2023-00445
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/25/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/19/2018
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-