• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM; EXTERNAL COMPONENT Back to Search Results
Model Number SONNET RECHARGEABLE BATTERY [OPUS 3 RB]
Device Problems Crack (1135); Fracture (1260)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
 
Event Description
A broken sonnet rcb has been received at hq.Upon preliminary investigation by service repair, there was no leaking electrolyte observed.
 
Manufacturer Narrative
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed, most likely due to external mechanical stress.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.Despite requested no further information could be retrieved from the field.This is a final report.
 
Event Description
A broken sonnet rcb has been received at hq.Upon investigation by service repair there was no leaking electrolyte observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MED-EL MAESTRO COCHLEAR IMPLANT SYSTEM
Type of Device
EXTERNAL COMPONENT
Manufacturer (Section D)
MED-EL ELEKTROMEDIZINISCHE GERAETE GMBH
innsbruck
AU 
Manufacturer Contact
laura simonotti
fuerstenweg 77a
innsbruck 6020
AU   6020
57788
MDR Report Key16251890
MDR Text Key308253840
Report Number9710014-2023-00062
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeAU
PMA/PMN Number
P000025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberSONNET RECHARGEABLE BATTERY [OPUS 3 RB]
Device Lot NumberSN: 2107N074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2023
Was the Report Sent to FDA? No
Date Manufacturer Received01/16/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-