Model Number SONNET RECHARGEABLE BATTERY [OPUS 3 RB] |
Device Problems
Crack (1135); Fracture (1260)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device has been returned to med-el hq where it will be evaluated.When available, a device failure analysis will be submitted as a follow-up report.
|
|
Event Description
|
A broken sonnet rcb has been received at hq.Upon preliminary investigation by service repair, there was no leaking electrolyte observed.
|
|
Manufacturer Narrative
|
Conclusion: the returned device was visually inspected upon arrival.A mechanically damaged housing was observed, most likely due to external mechanical stress.The observed damage did not allow electrical testing to be conducted.The damage to the battery housing was clearly the reason for the malfunction of the device returned by the end-user.Despite requested no further information could be retrieved from the field.This is a final report.
|
|
Event Description
|
A broken sonnet rcb has been received at hq.Upon investigation by service repair there was no leaking electrolyte observed.
|
|