• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 SITE~RITE 8 ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 SITE~RITE 8 ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number N/A
Device Problem Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/10/2023
Event Type  malfunction  
Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been received by the manufacturer for evaluation.A history review of serial number (b)(4) showed seven other similar product complaint(s) from this serial number.The similar complaints have been reported by the same uk facility.
 
Event Description
It was reported the device is not functioning correctly.It has been used to insert 8 picc lines all of which the device has claimed have been inserted and positioned correctly, how ever when the patients have been x-rayed the lines have been revealed to be incorrectly.This report addresses the eighth device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key16261245
MDR Text Key308841501
Report Number3006260740-2023-00196
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/28/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number9770550
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/13/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-