Model Number HEM1
Device Problem Sparking (2595)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2023
Event Type  malfunction  
Manufacturer Narrative
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product an investigation will be completed to consider any potential factors that may have contributed to this complaint and a supplemental report will be sent with the investigation results.
 
Event Description
As reported, with this demo hemosphere for edwards employee's, the power cable was split and sparks could be observed, with no smoke or fire.Edwards employee was not aware of liquid ingress to the cable.The issue did not cause any damage.Replacing the cable the problem was solved.There was no patient involved.Follow-up has started for device return.
 
Manufacturer Narrative
Update: d4 (expiration date), g3 (date received by manufacturer), h3 (device evaluated by manufacturer), h4 (manufacturer date), h6 (investigation findings, investigation conclusions) the damaged cable involved in this event was not available for evaluation since it was discarded.Therefore, a product non-conformance or device failure could not be confirmed.Without return of the product, edwards is unable to perform a complete investigation of the reported event.It is not possible to determine what factors may have contributed to it, and therefore no actions could be planned.Based of further engineering investigation the operator's manual was reviewed and it has multiple warnings regarding using the power cord with installed power cord entry cover.Failure to do so may result in fluid ingress.Also, not to use detachable power cords other than the power cord provided.The manual also warns: do not use any damaged or non-compatible platform accessories, components or cables.The manufacturing records were reviewed and there is no indication of a related nonconformance; all process parameters were met and inspections passed successfully.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.H3 other text : the cable was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEMOSPHERE ADVANCED MONITOR
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
samantha eveleigh
1 edwards way
irvine, CA 92614
9492503939
MDR Report Key16263395
MDR Text Key308964030
Report Number2015691-2023-10430
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K213682
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberHEM1
Device Catalogue NumberHEM1
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/13/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-