• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problems No Audible Prompt/Feedback (2282); Audible Prompt/Feedback Problem (4020)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
Date of event udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the fluid warmers speaker is not working.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Other, other text: h3.Device evaluated by manufacturer and h6.Event problem and evaluation codes: updated.H10 device evaluation: one (1) fluid warmer was received.A visual inspection noted a cracked tank cover, and wear and tear damaged enclosure and front cover.Performed functional testing.The speaker on the device didn't work confirming the customer complaint.The root cause was due to a faulty printed circuit board (pcb).A manufacturing device history record (dhr) review was not performed because there was no indication of a manufacturing defect during the investigation.Service history review identified this device was previously serviced for similar issues over a year ago in aug 2021 where the printed circuit board (pcb) was replaced.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
purchasing and stores
minneapolis, MN 55442
MDR Report Key16268909
MDR Text Key308563371
Report Number3012307300-2023-00731
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001764
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-90
Device Catalogue NumberCON-HL-90
Was Device Available for Evaluation? No
Date Returned to Manufacturer01/24/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-