Model Number HL-90 |
Device Problems
No Audible Prompt/Feedback (2282); Audible Prompt/Feedback Problem (4020)
|
Patient Problem
Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event udi section are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that the fluid warmers speaker is not working.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
Other, other text: h3.Device evaluated by manufacturer and h6.Event problem and evaluation codes: updated.H10 device evaluation: one (1) fluid warmer was received.A visual inspection noted a cracked tank cover, and wear and tear damaged enclosure and front cover.Performed functional testing.The speaker on the device didn't work confirming the customer complaint.The root cause was due to a faulty printed circuit board (pcb).A manufacturing device history record (dhr) review was not performed because there was no indication of a manufacturing defect during the investigation.Service history review identified this device was previously serviced for similar issues over a year ago in aug 2021 where the printed circuit board (pcb) was replaced.No action taken due to the age and condition of the device.It is deemed beyond economical repair and will be scrapped.
|
|