• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-2000; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-2000 ST
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2023
Event Type  malfunction  
Event Description
A customer reported error message ¿2250 sample loader rack x1-step feed failure¿ on the aia-2000 analyzer.The customer can load and eject the sample racks, but when rack is about to be sampled, the error reoccurs.The customer also reported no obstruction in the sample racks lane, performed a hard shut down of the analyzer and software, but the error persists.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Manufacturer Narrative
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and reproduced the complaint by performing rack movement test in the mainte function.While troubleshooting, fse found the x1 motor flag pitch out of alignment.Fse resolved the complaint by performing a x1 motor flag pitch adjustment.Fse validated the analyzer by successfully performing quality control run without error and within published package insert ranges.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for the serial number: (b)(4) from 06dec2021 through aware date 06jan2023.There were no similar complaints identified during the search period.Aia-2000 operator's manual on the appendix 4: error messages: (2250) sample loader rack x1-step feed failure cause: the pitch sensor did not activate during step feed (x1) operation.Solution: check the sample rack path for obstacles.If this error occurs frequently, contact tosoh service.The most probable cause of the reported event was due to misaligned loader x1 motor flag pitch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-2000
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key16269695
MDR Text Key309201444
Report Number3004529019-2023-00250
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189284616
UDI-Public04560189284616
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-2000 ST
Device Catalogue Number022100
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-