Model Number AIA-2000 ST |
Device Problem
Failure to Align (2522)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/06/2023 |
Event Type
malfunction
|
Event Description
|
A customer reported error message ¿2250 sample loader rack x1-step feed failure¿ on the aia-2000 analyzer.The customer can load and eject the sample racks, but when rack is about to be sampled, the error reoccurs.The customer also reported no obstruction in the sample racks lane, performed a hard shut down of the analyzer and software, but the error persists.The analyzer is down.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|
|
Manufacturer Narrative
|
A field service engineer (fse) was at the customer¿s site to address the reported event.Fse confirmed the complaint by reviewing the error logs and reproduced the complaint by performing rack movement test in the mainte function.While troubleshooting, fse found the x1 motor flag pitch out of alignment.Fse resolved the complaint by performing a x1 motor flag pitch adjustment.Fse validated the analyzer by successfully performing quality control run without error and within published package insert ranges.No further action required by field service.The aia-2000 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for the serial number: (b)(4) from 06dec2021 through aware date 06jan2023.There were no similar complaints identified during the search period.Aia-2000 operator's manual on the appendix 4: error messages: (2250) sample loader rack x1-step feed failure cause: the pitch sensor did not activate during step feed (x1) operation.Solution: check the sample rack path for obstacles.If this error occurs frequently, contact tosoh service.The most probable cause of the reported event was due to misaligned loader x1 motor flag pitch.
|
|
Search Alerts/Recalls
|
|
|