PHILIPS MEDICAL SYSTEMS MX40 2.4 GHZ SMART HOPPING Back to Search Results
Model Number 865351
Device Problems No Audible Alarm (1019); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information and the complaint is still under investigation.A follow-up report will be submitted upon completion of the investigation.(b)(6).
 
Event Description
The customer reported a speaker malfunction.It is unknown if the device was in clinical use at the time the issue was discovered.  there was no adverse event or patient harm reported.
 
Manufacturer Narrative
It is reported the device is in transit to the repair facility.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
Additional information was received which confirmed the device displayed a speaker malfunction message and had no audible sound.The device was reported to be in use on a patient, but no adverse event to the patient or user was reported.
 
Manufacturer Narrative
The customer was provided a replacement device.Attempts were made to have the complaint device returned for evaluation; however, there was no response, and the device has not been returned.
 
Manufacturer Narrative
H4 device manufacture date was corrected from (b)(6) 2016 to (b)(6) 2016.The device was returned to the philips authorized repair facility.Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that the speaker produced no sound and the speaker was defective.Based on the information available and the testing conducted, the cause of the reported problem was a defective speaker.The customer was previously provided a replacement device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 2.4 GHZ SMART HOPPING
Type of Device
MX40 2.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16270408
MDR Text Key308755011
Report Number1218950-2023-00055
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838082243
UDI-Public00884838082243
Combination Product (y/n)N
Reporter Country CodeRP
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number865351
Device Catalogue Number865351
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-