Model Number 0998-00-0800-53
Device Problems Detachment of Device or Device Component (2907); Optical Problem (3001); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2021
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted as the result of a retrospective review conducted in capa 584165.A supplemental report will be submitted when the evaluation is completed.
 
Event Description
It was reported that during a service repair performed by a getinge field service engineer (fse), the doppler tray hanging on just one hinge.There was no patient involvement.Add related complaint (b)(4).
 
Manufacturer Narrative
Updated fields: (site country), (medical device ¿ problem code, type of investigation, investigation findings, component codes, investigation conclusions).Corrected fields: (initial reporter: (b)(6) ) additional information: (contact spoc: (b)(6) ) during a repair service, the getinge field service engineer (fse) discovered that the doppler tray was hanging on just one hinge, to fix the issue he replaced the replaced the bin chassis.He then performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16271286
MDR Text Key309062391
Report Number2249723-2023-00825
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2015
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-