Manufacturer's investigation conclusion: the reported event of the patient not hearing alarms was not confirmed.It was reported that when the patient presented in clinic, the healthcare professional noticed a pump stop a few days prior to the clinic visit in addition to low voltage alarms in the controller history.The patient reported not having any alarms or noticing anything being different.The controller was reported to be exchanged and subsequently discarded.No product was returned for evaluation and log files from the exchanged controller were not provided.The root cause of the reported event could not be conclusively determined through this analysis.Device history records were reviewed and showed no deviations from manufacturing or quality assurance (qa) specifications.Heartmate ii patient handbook (rev.C), under section 5 ¿alarms and troubleshooting¿ and heartmate ii instructions for use (ifu) (rev.B), under section 7 ¿alarms and troubleshooting¿ cover all alarms (visual and audible), including the pump off and low voltage hazard alarm conditions, and the actions to take if the alarms cannot be resolved.Heartmate ii patient handbook (rev.C) section 6 "caring for the equipment" and heartmate ii ifu (rev.B) section 8 "equipment storage and care" describe how to care for and clean all equipment.This section instructs the user to, at least monthly, inspect the system controller¿s audio sounders for dirt or grease.If a change in tone or in loudness is noticed during a system controller self-test, the audio speaker sockets may be obstructed.Audio speaker sockets may be cleaned using a small cotton swab that is moistened (not dripping) with rubbing alcohol.Never insert anything sharp (like a toothpick or pin) into the sounder holes.This can damage the speakers inside.Heartmate ii patient handbook (doc #10006962, rev.C) section 2 "how your heart pump works" and heartmate ii instructions for use (ifu) (doc #10006960, rev.C) section 2 "system operations" explain that the system controller self test should be performed at least once per day to check the audible and visual alarm indicators on the user interface, as well as the status of the backup battery.Heartmate ii patient handbook (rev.C) section 10 and heartmate ii ifu (rev.B) section f, both entitled ¿safety checklists¿, provide checklists to assist the patient in performing routine maintenance of heartmate 3 lvad.This section also informs the user to replace any equipment or system component that appears damaged or worn.Heartmate ii patient handbook (rev.C) and heartmate ii ifu (rev.B) under section 3, entitled ¿powering the system¿, state that the patient must always connect to the mobile power unit or power module for sleeping or when there is a chance of sleep, as a sleeping patient may not hear the system controller alarms.Heartmate ii patient handbook (rev.C) cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|