• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR DRAGONFLY¿ OPTIS¿ IMAGING CATHETER; DIAGNOSTIC INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR DRAGONFLY¿ OPTIS¿ IMAGING CATHETER; DIAGNOSTIC INTRAVASCULAR CATHETER Back to Search Results
Model Number C408646
Device Problem Gas/Air Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2023
Event Type  malfunction  
Event Description
It was reported that during flushing, an air bubble appeared at the distal tip of the dragonfly optis imaging catheter.There was no patient involvement.No additional information was provided.
 
Manufacturer Narrative
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Manufacturer Narrative
Visual and functional analysis was performed on the returned device.The reported gas/air leak was unable to be confirmed due to the condition of the returned unit; however, the window tube was noted to be stretched/wrinkled and ballooned.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar complaints reported from this lot.Based on the reported information and analysis of the returned unit, the noted stretched/wrinkled and ballooned window tube appears to be related to a potential product quality issue; therefore, an exception issue has been initiated to further investigate this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ IMAGING CATHETER
Type of Device
DIAGNOSTIC INTRAVASCULAR CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT MEDICAL REG#3009600098
4 robbins rd
westford MA 01886
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key16286294
MDR Text Key309014095
Report Number2024168-2023-01106
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K141769 
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC408646
Device Catalogue NumberC408646
Device Lot Number8677039
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/13/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-