SYNTHES GMBH RADIAL HEAD W/ Ø22MM HD, Ø6.5MM STM,10(+3)MM HD HT,STER; PROSTHESIS, ELBOW, HEMI-, RADIAL, POLYMER
|
Back to Search Results |
|
Catalog Number 09.405.263S |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Unspecified Tissue Injury (4559)
|
Event Type
Injury
|
Event Description
|
Device report from depuy synthes reports an event in switzerland as follows: it was reported that a year after implantation of the radial head replacement system (rhrs), the x-rays taken on an unknown date showed a light border (lysboarder) around the stem.Patient outcome is reported as happy and in good health.No further information provided.This report is for one (1) radial head w/ ø22mm hd, ø6.5mm stm,10(+3) mm hd ht,ster.This is report 1 of 1 for complaint (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
No other medical intervention was required.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H6: the device lot number is unknown, therefore a device history review could not be performed.If more information become available, the record will be re-assessed.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the device from the received image(s).Visual analysis of the photo revealed that there was no damage or defects with the rad head w/head ø22 shaft ø6.5 hh10 +3.There is not enough evidence to confirm the reported allegation based on the pictures provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the rad head w/head ø22 shaft ø6.5 hh10 +3 was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|
|