Model Number ALLURA XPER FD20
Device Problems Appropriate Term/Code Not Available (3191); Radiation Output Failure (4027)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Event Description
It has been reported to philips that wireless footswitch was defective.No harm has been reported to philips.We are conservatively reporting this event as the investigation is ongoing.A follow up report will be submitted when further information is received.
 
Manufacturer Narrative
Philips has investigated this case.According to the information collected, the customer inquired about the wireless footswitch replacement status for a previously reported defective wireless footswitch, which was addressed in complaint tw 1996801.Philips informed the customer regarding the availability of the replacement wireless footswitch and the service case was closed.The customer continued to use a wired footswitch until the replacement footswitch was later provided (tw 3640240).The system has been returned to use in good working order.At the time this complaint was received, philips did not have enough information to exclude the potential for death or serious injury on recurrence, and as such the complaint was reported.Since that time, new information has been received which has led to the determination that this is scenario is not likely to cause or contribute to death or serious injury if it were to recur.There was no reported product malfunction in this case.The customer had requested information regarding the availability of a replacement wireless footswitch.Philips has re-evaluated this case as not reportable.Codes were updated based on investigation results.Device problem code was corrected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD20
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16297277
MDR Text Key308837359
Report Number3003768277-2023-00726
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K033737
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722006
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/18/2008
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-