Catalog Number 67-2566
Device Problem Unable to Obtain Readings (1516)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/12/2023
Event Type  malfunction  
Manufacturer Narrative
Udi information is unknown.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event Description
It was reported that the lcd pot was coming off board leading to it not being read correctly.No patient involvement.
 
Manufacturer Narrative
Device manufacture date, d10.Device available for evaluation, h3.Device evaluated by manufacturer, h6.Health impact and evaluation codes: updated.D3, g1,2 email is: (b)(6.One printed circuit board (pcb) was received in with a disconnected potentiometer.Visual inspection confirmed the complaint.The root cause was a broken pot off pcb board from impact during shipping and handling.A review of the product history record indicated that the kitted board was fully inspected and showed no physical damage before packaging per assembly process.No action taken, as a replacement was sent to the reporting customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE BLOOD AND FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
3350 granada ave n
oakdale MN 55128
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16297944
MDR Text Key308850823
Report Number3012307300-2023-00876
Device Sequence Number1
Product Code LGZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number67-2566
Device Lot Number4342238
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-