The graft bolt is a component of the alphatec identiti alif standalone interbody system, which is an integrated intervertebral body fusion device for use in anterior lumbar interbody fusion (alif) procedures used in patients for treatment, not diagnosis.Photographs/radiographs were not provided.The surgery date is unknown.The patient's bone quality and medical history are unknown.No additional information has been provided to alpahtec regarding this event.The distal portion of the graft bolt did not return to alphatec.The proximal portion of the broken graft bolt was returned to alphatec on 01/04/2023.Visual inspection of the graft bolt indicates thread damage to the medial portion of the bolt.This type of failure can occur during insertion if there is shallow or incorrect graft bolt trajectory which can cause it to impinge on the posterior end of the medial screw hole or posterior wall of the interbody spacer.This could lead to an increase in torque by the user resulting in the shear failure.A good faith effort is being made to obtain the applicable information relevant to the report.If additional information is obtained, a supplemental report will be submitted.Labeling review: "warnings/cautions/precautions: care should be taken in performing screw hole preparation to facilitate a proper graft bolt insertion trajectory and implantation.Confirm under fluoroscopy that the graft bolt insertion angle is as close as possible to a 40° trajectory.A shallow or incorrect graft bolt trajectory may result in encroachment of the spacer and lead to graft bolt breakage.".
|