Model Number 232-10-85-300
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The graft bolt is a component of the alphatec identiti alif standalone interbody system, which is an integrated intervertebral body fusion device for use in anterior lumbar interbody fusion (alif) procedures used in patients for treatment, not diagnosis.Photographs/radiographs were not provided.The surgery date is unknown.The patient's bone quality and medical history are unknown.No additional information has been provided to alpahtec regarding this event.The distal portion of the graft bolt did not return to alphatec.The proximal portion of the broken graft bolt was returned to alphatec on 01/04/2023.Visual inspection of the graft bolt indicates thread damage to the medial portion of the bolt.This type of failure can occur during insertion if there is shallow or incorrect graft bolt trajectory which can cause it to impinge on the posterior end of the medial screw hole or posterior wall of the interbody spacer.This could lead to an increase in torque by the user resulting in the shear failure.A good faith effort is being made to obtain the applicable information relevant to the report.If additional information is obtained, a supplemental report will be submitted.Labeling review: "warnings/cautions/precautions: care should be taken in performing screw hole preparation to facilitate a proper graft bolt insertion trajectory and implantation.Confirm under fluoroscopy that the graft bolt insertion angle is as close as possible to a 40° trajectory.A shallow or incorrect graft bolt trajectory may result in encroachment of the spacer and lead to graft bolt breakage.".
 
Event Description
Alphatec received information indicating a graft bolt broke during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IDENTITI ALIF STANDALONE INTERBODY SYSTEM
Type of Device
INTERVERTEBRAL FUSION DEVICE WITH INTEGRATED FIXATION, LUMBAR
Manufacturer (Section D)
ALPHATEC SPINE, INC.
1950 camino vide roble
carlsbad CA 92008
Manufacturer (Section G)
ALPHATEC SPINE, INC.
1950 camino vide roble
carlsbad CA 92008
Manufacturer Contact
wesley channell
1950 camino vide roble
carlsbad, CA 92008
9014283693
MDR Report Key16301451
MDR Text Key308871013
Report Number2027467-2023-00003
Device Sequence Number1
Product Code OVD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K203742
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number232-10-85-300
Device Catalogue Number232-10-85-300
Device Lot NumberTO00835
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2023
Date Manufacturer Received01/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-